#PAGE_PARAMS# #ADS_HEAD_SCRIPTS# #MICRODATA#

Intubation with channeled versus non-channeled video laryngoscopes in simulated difficult airway by junior doctors in an out-of-hospital setting: A crossover manikin study


Authors: Shi Hao Chew aff001;  Jonathan Zhao Min Lim aff001;  Benjamin Zhao Bin Chin aff001;  Jia Xin Chan aff001;  Raymond Chern Hwee Siew aff003
Authors place of work: Department of Anaesthesia, National University Hospital, Singapore aff001;  Headquarters Army Medical Services, Singapore Armed Forces, Singapore aff002;  RS Anaesthesia & Intensive Care, Singapore aff003
Published in the journal: PLoS ONE 14(10)
Category: Research Article
doi: https://doi.org/10.1371/journal.pone.0224017

Summary

Failure to secure the airway is an important cause of morbidity and mortality during resuscitations. We compared the rate of successful intubation of the King Vision aBlade channeled and non-channeled video laryngoscopes, and McGRATH MAC video laryngoscope when used by junior doctors to intubate a simulated difficult airway in an out-of-hospital setting. 105 junior doctors were recruited in a crossover study to perform tracheal intubation with the three video laryngoscopes on a simulated difficult airway using the SimMan® 3G manikin. Primary outcome was the rate of successful intubations. Secondary outcomes were time-to-visualization, time-to-intubation and ease of use. Rates of successful intubations were higher for King Vision channeled and McGrath compared to the King Vision non-channeled (85.7% and 82.9% respectively versus 24.8%; p<0.001). Amongst the participants who had successful intubations, King Vision channeled and McGrath had shorter mean time-to-intubation compared to the King Vision non-channeled (41.3±20.3s and 38.5±18.7s respectively versus 53.8±23.8s, p<0.004;). There was no significant difference in the rate of successful intubation and mean time-to-intubation between King Vision channeled and McGrath. The King Vision channeled and McGrath video laryngoscopes demonstrated superior intubation success rates compared to King Vision non-channeled laryngoscope when used by junior doctors for intubating simulated difficult airway in an out-of-hospital setting. We postulated that the presence of a guidance channel in the King Vision channeled laryngoscope and the familiarity of the blade curvature and handling of the McGrath could have accounted for their improved intubation success rates.

Keywords:

Resuscitation – Medical doctors – Face recognition – morbidity – Intubation – Tongue – Pilot studies

Introduction

Failure to secure the airway is an important cause of morbidity and mortality during resuscitations [1,2]. Video laryngoscopes have been consistently demonstrated to be critical in difficult or failed intubations [35] and have been incorporated into many difficult airway algorithms [68].

The earlier generation of video laryngoscopes were expensive, bulky, and required familiarization training [9]. This precluded widespread use of video laryngoscopes apart from specialised areas such as operating rooms and intensive care units [10]. Newer video laryngoscopes are now more portable, with some models incorporating additional features such as the channeled blade. The presence of this channel can help to reduce the need for complex manipulation of the tracheal tube during intubation [1112]. Although there are many studies comparing the intubation efficacy of different video laryngoscopes, few studies have compared the success rate and ease of use of the channeled versus non-channeled video laryngoscopes when used by junior doctors in an out-of-hospital or military setting [1214].

Our primary aim was to compare the rate of successful intubation of the channeled and non-channeled blade video laryngoscopes when used by junior doctors to intubate a simulated difficult airway. Secondary outcomes assessed included time-to-visualization of glottis, time-to-intubation and ease of use of the video laryngoscopes. We also investigated the impact of blade curvature of the non-channeled video laryngoscope blades on their intubating efficacies.

Materials and methods

A crossover study was conducted from January 2016 to March 2017. Study was approved by local institutional research committee SAF Joint Medical Conference (Research) BB52/12-4 on 2 Aug 2015. Written consent was obtained from participants.

All junior doctors from the Medical Officer Cadet Course of the Singapore Armed Forces were considered for this study. Junior doctors were defined as post graduate year two or three medical officers with 10 or less previous intubation attempts in real-life patients. Exclusion criteria were medical officers with more than 10 previous intubation attempts in real-life patients, refusal to give written informed consent or non-physician medical officers.

In this study, we used the acute angled channeled and non-channeled blades of King Vision aBlade (King Systems, Noblesville, IN, USA) video laryngoscope, and the Macintosh styled blade of the McGRATH MAC (Aircraft Medical, Edinburgh, UK) video laryngoscope. We chose the King Vision video laryngoscope because of its cost, portability and design for use in patients with limited mouth opening (≥18mm and ≥13mm for the size 3 channeled and non-channeled blades respectively [15]) and restricted neck movement. These are important considerations in the management of difficult airway in an out-of-hospital or military setting. We also compared King Vision non-channeled with McGrath video laryngoscopes as McGrath uses a Macintosh blade design with a curvature that is less angled than that of King Vision.

All participants had to watch a standardised training video on the laryngoscopy technique using all three video laryngoscopes. We played the training video for KingVision (02m:05s) first, before the Mcgrath teaching video (00m:25s). The participants then undergo a familiarization session with the SimMan® 3G advanced patient simulator manikin prior to the start of the simulation session. To minimize the effects of time and fatigue, all the sessions took place during office hours, and none of the participants were post-call/ sleep-fatigued when participating in the study.

The participants intubated with each of the three devices following a sequence based on a 3-period, 3-treatment crossover design (Table 1) to minimise the learning effect on subsequent intubation attempts [11,14,16]. All participants were allowed up to 90 seconds to achieve tracheal intubation in a standardised simulated difficult airway scenario for each device. In the event of an oesophageal intubation, the timer continued and participants were allowed repeated attempts at intubation, up to a maximum of 90 seconds.

Tab. 1. Intubation sequence based on a 3-period, 3-treatment crossover design.
Intubation sequence based on a 3-period, 3-treatment crossover design.

Successful intubation was defined as the ability to achieve tracheal intubation within 90 seconds and was demonstrated with the inflation of the manikin’s lungs with a manual resuscitator. Time-to-visualization was defined as the time from taking hold of the laryngoscope handle to the visualization of the glottis by the participant. Time-to-intubation was defined as the total time taken from holding the laryngoscope handle to the confirmation of tracheal tube position by inflation of the manikin’s lungs. Upon completion of the simulation session, participants were asked to indicate their preferred device in an outfield setting and to rate the ease of use of the video laryngoscopes using a Likert 5-point scale (1 = very easy, 2 = easy, 3 = moderate, 4 = difficult, 5 = very difficult) in an anonymous survey.

A consistent difficult airway scenario was simulated by activating the tongue-oedema feature in the SimMan® 3G manikin and by placing a standard cervical collar for cervical immobilisation (Fig 1). The participants also had to intubate the manikin on the ground, simulating intubation in an out-of-hospital setting. All tracheal intubations were performed using the size 3 blade for each of the three video laryngoscopes and a standard size 7 Portex® tracheal tube in the SimMan® manikin. A pre-formed stylet was inserted into the Portex® tracheal tube for intubation with the King Vision non-channeled and McGrath laryngoscopes, while no stylet was used with King Vision channeled laryngoscopes.

Fig. 1. (A) Demonstration on the use of the King Vision laryngoscope and (B) visualization of vocal cords in the SimMan® 3G manikin during the difficult airway scenario.
(A) Demonstration on the use of the King Vision<sup>™</sup> laryngoscope and (B) visualization of vocal cords in the SimMan<sup>®</sup> 3G manikin during the difficult airway scenario.

A pilot study conducted earlier with 21 junior doctors showed a rate of successful intubation of 85% when using the McGrath in the SimMan® simulated difficult airway scenario. A power analysis was done to determine sample size based on this pilot study. We hypothesised that the King Vision channeled video laryngoscope will improve rate of successful intubation for junior doctors by 10% in the SimMan® simulated difficult airway scenario. Using a two-sided alpha level of 5% and statistical power of 80%, the required sample size to detect a 10% difference in the rate of successful intubation between the groups was calculated to be 77.

Statistical analysis was performed using SPSS Version 24 (SPSS Inc, Chicago IL, USA). The normal distribution of data was tested using the Shapiro-Wilk test. Comparison of success rates was analysed using Chi-squared tests. Analyses of continuous data were performed using one-way ANOVA (for parametric data) and independent-samples Kruskal-Wallis test (for non-parametric data) with Bonferroni correction. The ease of use of video laryngoscopes was analysed using Kruskal-Wallis test. A p-value of < 0.05 was considered significant.

Results

A total of 105 junior doctors were recruited in this study. The median number of successful real-life intubations performed based on recall amongst participants was three. None of the participants had used the King Vision and McGrath laryngoscopes for real-life intubations prior to this study.

Rates of successful intubations were higher for the King Vision channeled and McGrath compared to the King Vision non-channeled video laryngoscopes (85.7% and 82.9% respectively versus 24.8%; p<0.001, Table 2). However, there was no significant difference in the rates of successful intubations between the King Vision channeled and McGrath. All participants managed to achieve visualization of the glottis, with a mean time-to-visualization of less than 15 seconds for all three devices.

Tab. 2. Comparison of successful intubation and time-to-visualization between the King Vision channeled and non-channeled groups, and McGrath.
Comparison of successful intubation and time-to-visualization between the King Vision channeled and non-channeled groups, and McGrath.
Values are number (proportion) or mean (standard deviation).

Amongst participants who had successful intubations, King Vision channeled and McGrath had shorter mean time-to-intubation compared to the King Vision non-channeled (41.3±20.3s and 38.5±18.7s respectively versus 53.8±23.8s, p<0.004; Table 3). There was no significant difference in mean time-to-intubation between King Vision channeled and McGrath.

Tab. 3. Comparison of time-to-intubation for King Vision channeled and non-channeled groups, and McGrath.
Comparison of time-to-intubation for King Vision channeled and non-channeled groups, and McGrath.
Results are expressed as mean (standard deviation).

On a Likert 5-point scale for ease of use of video laryngoscopes, the median score for King Vision channeled and McGrath groups were 2 [interquartile range of 1–3], while the median score for King Vision non-channeled group was 4 [interquartile range of 3–5]. When asked to recommend a video laryngoscope amongst the three, 45.2% of participants recommended King Vision channeled, 1.5% recommended King Vision non-channeled, and 53.3% recommended McGrath.

Discussion

Video laryngoscopes are designed to provide indirect visualization of the glottis. Acute angled blades, such as the ones in King Vision video laryngoscopes, enable the user to achieve better laryngeal view with less neck and laryngeal manipulation [8,17]. However, without a direct line of vision to the glottis, the passage of tracheal tube into the glottis require deft hand-eye coordination [9,17], which may pose a problem for junior doctors with limited experience in intubation [1719]. The presence of a tracheal tube guidance channel can help to overcome the need for complex manipulation during intubation [1112]. Our study showed that the King Vision channeled laryngoscopes had better success rates compared to the King Vision non-channeled laryngoscopes, despite having the same blade angulation and curvature. This is consistent with the study by Akihisa et al that having a guided channel for tracheal tube improved intubation performance compared to a non-channeled blade for novice operators [14].

In our study, we made the intubation process more realistic by simulating a difficult airway with tongue oedema, cervical immobilisation with standard cervical collar, and requiring participants to intubate the manikin on the ground, similar to an out-of-hospital or military setting. The importance of a guided channel for an acute angled blade is accentuated when intubation becomes more challenging. This was demonstrated in our study where the success rate for King Vision non-channeled was 24.8%, compared to 47.3% for King Vision non-channeled in the study by Akihisa et al, where a standard manikin airway was used. This observation was also echoed in a study by Okada et al, which yielded no significant difference in intubation success rates between the King Vision channeled and non-channeled groups when used by novice operators in standard manikin airways, but demonstrated higher intubation success rates with the King Vision channeled laryngoscopes when compared to the non-channeled ones (25/25 versus 18/25, p = 0.004) when intubation was performed on a manikin with ongoing chest-compressions [12].

McGrath also performed better than the King Vision non-channeled laryngoscopes in terms of intubation success rates. While the McGrath did not have a channeled blade, it had the advantage of a blade curvature similar to the standard Macintosh blade, which the junior doctors were more familiar with. Other studies had also observed that junior doctors were able to translate their intubating skills if the curvature of the video laryngoscope blade was similar to the standard Macintosh blade [2021]. We postulated that it was possible that our participants found it easier to acquaint themselves with the McGrath and achieved better success rates than with the King Vision non-channeled laryngoscope because of their familiarity with the McGrath blade curvature and handling.

When designing our study, we believe that it was important to break down the intubation process into two components: (1) achieving glottis visualization followed by (2) manipulation of tracheal tube past the glottis. Many studies agreed that the improved laryngeal view associated with video laryngoscope use did not necessarily lead to intubation success [3,13,2122]. In our study, all 3 video laryngoscopes achieved a mean time-to-visualization of less than 15s. However, the King Vision non-channeled laryngoscope had the longest time-to-intubation. To improve intubation success rates with the King Vision non-channeled laryngoscope, we postulated that operators should be well-trained in the use of stylets, which some considered to be mandatory for intubation with angulated blades without a guiding channel [17,23]. They should be adept at shaping the stylets optimally for intubation, and possess good hand-eye coordination for manipulation of the tracheal tube when the oro-pharyngeal axis and the pharyngo-glotto-tracheal axis are not aligned to provide direct glottic view [9,13,17]. It will be interesting to determine whether the time to intubation using video laryngoscopes with acute angulated blades without a guided channel, such as the King Vision non-channeled video laryngoscope, differs between expert video laryngoscopists and less experienced medical personnel when intubating difficult airways.

There was no significant difference in terms of intubation success rates between the King Vision channeled and McGrath laryngoscopes. There were also no significant differences between King Vision channeled and McGrath in terms of time-to-visualization and time-to-intubation. Participants also rated both King Vision channeled and McGrath to be “easy to use”.

Other studies comparing various video laryngoscopes have also found that a variety of video laryngoscope are potentially appropriate for novice operators [4,11,13,24]. McGrath, with its standard Macintosh blade, offered the advantage of improved laryngeal view for operators who are familiar with its use. In a randomised controlled trial done by B. Alvis et al, experienced laryngoscopists with limited experience in video laryngoscopes were more successful and intubated faster with McGrath compared to King Vision channeled laryngoscopes [20]. On the other hand, novice operators, who might not be as adept with the technique of direct laryngoscopy, might prefer the King Vision channeled laryngoscopes because of its easy manipulation past the tongue for glottis visualization and a channel for tracheal guidance into the glottis [2528]. King Vision channeled and non-channeled laryngoscopes have also been described for use as an alternative to awake fibreoptic intubation in patients with limited mouth opening and/or neck movement [15,29].

There were several limitations to our study. Firstly, while the SimMan® manikins used in this study are high fidelity advanced patient simulators, intubating a manikin is different from intubating a real patient. Hence the results might not be directly applicable to real-life patients. Secondly, while the SimMan® allowed for standardisation of the difficult airway scenario for our participants, this was limited to cervical immobilisation and tongue oedema. We recognised that difficult airways scenarios in actual trauma or collapsed patients may include other variables such as blood and secretions in the oral cavity, anatomic variations and/or facial injuries, which are beyond the scope of this study. Demographic differences, which were not measured, could have made a difference to the ease of intubation. Taller, heavier participants might find it difficult to kneel down on the ground to intubate compared to smaller participants.

Supporting information

S1 Fig [png]

S1 File [sav]
Data Set.

S1 Table [docx]

S2 Table [docx]

S3 Table [docx]


Zdroje

1. Mort TC. Emergency tracheal intubation: complications associated with repeated laryngoscopic attempts. Anesthesia & Analgesia 2004; 99: 607–613.

2. Martin LD, Mhyre JM, Shanks AM, Tremper KK, Kheterpal S. 3,423 emergency tracheal intubations at a university hospital: airway outcomes and complications. Anesthesiology 2011; 114: 42–48. doi: 10.1097/ALN.0b013e318201c415 21150574

3. Vaida S, Budde A. Is it time for videolaryngoscopy to be the standard of care? Minerva Anestesiologica 2016; 82: 1–2. 26222394

4. Van Zundert A, Pieters B. Videolaryngoscopy: the new standard for intubation. Ten years’ experience. Minerva Anestesiologica 2015; 81: 1159–1162. 26339751

5. Piepho T, Fortmueller K, Heid FM, Schmidtmann I, Werner C, Noppens RR. Performance of the C-MAC video laryngoscope in patients after a limited glottic view using Macintosh laryngoscopy. Anaesthesia 2011; 66: 1101–1105. doi: 10.1111/j.1365-2044.2011.06872.x 21883131

6. Apfelbaum JL, Hagberg CA, Caplan RA, Blitt CD, Connis RT, Nickinovich DG. Practice guidelines for management of the difficult airway: an updated report by the American Society of Anesthesiologists Task Force on Management of the Difficult Airway. Anesthesiology 2013; 118: 251–270. doi: 10.1097/ALN.0b013e31827773b2 23364566

7. Frerk C, Mitchell VS, McNarry AF, Mendonca C, Bhagrath R, Patel A, et al. Difficult Airway Society 2015 guidelines for management of unanticipated difficult intubation in adults. British Journal of Anaesthesia 2015; 115: 827–848. doi: 10.1093/bja/aev371 26556848

8. Law JA, Broemling N, Cooper RM, Drolet P, Duggan LV, Griesdale DE, et al. The difficult airway with recommendations for management—Part 1 –difficult tracheal intubation encountered in an unconscious/induced patient. Canadian Journal of Anesthesia 2013; 60: 1089–1118. doi: 10.1007/s12630-013-0019-3 24132407

9. Cooper RM, Pacey JA, Bishop MJ, McCluskey SA. Early clinical experience with a new videolaryngoscope (GlideScope) in 728 patients. Canadian Journal of Anaesthesia 2005; 52: 191–198. 15684262

10. Tang KC, Chiu JW, Low E. Airway and ventilatory equipment in field anesthesia: what’s new? Military Medicine 2004; 169: 342–348. doi: 10.7205/milmed.169.5.342 15185996

11. Pieters BM, Wilbers NE, Huijzer M, Winkens B, van Zundert AA. Comparison of seven videolaryngoscopes with the Macintosh laryngoscope in manikins by experienced and novice personnel. Anaesthesia 2016; 71: 556–564. doi: 10.1111/anae.13413 26973253

12. Okada D, Komasawa N, Fujiwara S, Minami T. Comparison of tube-guided and guideless videolaryngoscope for tracheal intubation during chest compression in a manikin: a randomized crossover trial. Journal of Anesthesia 2015; 29: 331–337. doi: 10.1007/s00540-014-1936-1 25348686

13. Kleine-Brueggeney M, Greif R, Schoettker P, Savoldelli GL, Nabecker S, Theiler LG. Evaluation of six videolaryngoscopes in 720 patients with a simulated difficult airway: a multicentre randomized controlled trial. British Journal of Anaesthesia 2016; 116: 670–679. doi: 10.1093/bja/aew058 27106971

14. Akihisa Y, Maruyama K, Koyama Y, Yamada R, Ogura A, Andoh T. Comparison of intubation performance between the King Vision and Macintosh laryngoscopes in novice personnel: a randomized, crossover manikin study. Journal of Anesthesia. 2014; 28: 51–57. doi: 10.1007/s00540-013-1666-9 23812581

15. de Pinho Martins M, Bastos AM, Fontes CL, Pinto Calcada ABC. Case report: tracheal intubation with King Vision in a patient with oral opening <1cm. European Journal of Anaesthesiology 2014; 31: 270.

16. Foulds LT, McGuire BE, Shippey BJ. A randomised cross-over trial comparing the McGrath® Series 5 videolaryngoscope with the Macintosh laryngoscope in patients with cervical spine immobilisation. Anaesthesia 2016; 71: 437–442. doi: 10.1111/anae.13384 26849094

17. Levitan RM, Heitz JW, Sweeney M, Cooper RM. The complexities of tracheal intubation with direct laryngoscopy and alternative intubation devices. Annals of Emergency Medidine 2011; 57: 240–247.

18. Randell T. Prediction of difficult intubation. Acta Anaesthesiologica Scandinavica 1996; 40: 1016–1023. doi: 10.1111/j.1399-6576.1996.tb05620.x 8908216

19. Orebaugh SL. Difficult airway management in the emergency department. The Journal of Emergency Medicine 2002; 22: 31–48. doi: 10.1016/s0736-4679(01)00435-8 11809554

20. Alvis BD, Hester D, Watson D, Higgins M, St Jacques P. Randomized controlled trial comparing the McGrath MAC video laryngoscope with the King Vision video laryngoscope in adult patients. Minerva Anestesiologica 2016; 82: 30–35. 25881731

21. Herbstreit F, Fassbender P, Haberl H, Kehren C, Peters J. Learning endotracheal intubation using a novel videolaryngoscope improves intubation skills of medical students. Anesthesia & Analgesia 2011; 113: 586–590.

22. Lewis SR, Butler AR, Parker J, Cook TM, Smith AF. Videolaryngoscopy versus direct laryngoscopy for adult patients requiring tracheal intubation. Cochrane Database of Systematic Reviews 2016; 11: CD011136. doi: 10.1002/14651858.CD011136.pub2 27844477

23. Gawlowski P, Iskrzycki L. Comparison of the KingVision video laryngoscope to the Macintosh laryngoscope for intubation of patient with cervical spine immobilization using extrication collars. The American Journal of Emergency Medicine 2017; 35: 657–658. doi: 10.1016/j.ajem.2016.12.064 28057379

24. Lewis SR, Butler AR, Parker J, Cook TM, Schofield-Robinson OJ, Smith AF. Videolaryngoscopy versus direct laryngoscopy for adult patients requiring tracheal intubation: a Cochrane Systematic Review. British Journal of Anaesthesia 2017; 119: 369–383. doi: 10.1093/bja/aex228 28969318

25. Escott MEA, Gleisberg GR, Gillum L, Cosper J, Traynor KM, Aulbert L, et al. Deploying the video laryngoscope into a ground EMS system. Journal of Emergency Medical Services 2014; August: 34–39

26. Gaszynska E, Samsel P, Stankiewicz-Rudnicki M, Wieczorek A, Gaszynski T. Intubation by paramedics using the ILMA or AirTraq, King Vision, and Macintosh laryngoscopes in vehicle-entrapped patients: a manikin study. European Journal of Emergency Medicine 2014; 21: 61–64. 23778270

27. Gawlowski P, Smereka J, Madziala M, Cohen B, Ruetzler K, Szarpak L. Comparison of the ETView Single Lumen and Macintosh laryngoscopes for endotracheal intubation in an airway manikin with immobilized cervical spine by novice paramedics: A randomized crossover manikin trial. Medicine 2017; 96: 1–6.

28. Szarpak Ł, Truszewski Z, Smereka J, Evrin T, Czyzewski Ł. Are paramedics able to perform endotracheal intubation with access to the patient through the back seat of the car? Randomized crossover manikin study. The American Journal of Emergency Medicine 2016; 34: 1161–1163. doi: 10.1016/j.ajem.2016.02.062 27079500

29. Gaszynska E, Gaszynski T. The King Vision™ video laryngoscope for awake intubation: series of cases and literature review. Therapeutics and Clinical Risk Management 2014; 10: 475–478. doi: 10.2147/TCRM.S64638 25018634


Článek vyšel v časopise

PLOS One


2019 Číslo 10
Nejčtenější tento týden
Nejčtenější v tomto čísle
Kurzy

Zvyšte si kvalifikaci online z pohodlí domova

plice
INSIGHTS from European Respiratory Congress
nový kurz

Současné pohledy na riziko v parodontologii
Autoři: MUDr. Ladislav Korábek, CSc., MBA

Svět praktické medicíny 3/2024 (znalostní test z časopisu)

Kardiologické projevy hypereozinofilií
Autoři: prof. MUDr. Petr Němec, Ph.D.

Střevní příprava před kolonoskopií
Autoři: MUDr. Klára Kmochová, Ph.D.

Všechny kurzy
Kurzy Podcasty Doporučená témata Časopisy
Přihlášení
Zapomenuté heslo

Zadejte e-mailovou adresu, se kterou jste vytvářel(a) účet, budou Vám na ni zaslány informace k nastavení nového hesla.

Přihlášení

Nemáte účet?  Registrujte se

#ADS_BOTTOM_SCRIPTS#