“Snags” connected with establishing the vitamin D
Authors:
Anton Vaňuga 1,2; Peter Huba 1; Ivana Ságová 3; Pavel Blažíček 1; Jarmila Melegová 5; Juraj Payer 4; Peter; Vaňuga 2
Authors place of work:
Alphamedical s. r. o., Slovenská republika
1; Národný endokrinologický a diabetologický ústav, n. o., Ľubochňa, Slovenská republika
2; I. interná klinika JLF UK a UNM, Martin, Slovenská republika
3; V. interná klinika LF UK a UNB, Nemocnica Ružinov, Bratislava, Slovenská republika
4; Klinická biochémia, s. r. o, Žilina, Slovenská republika
5
Published in the journal:
Vnitř Lék 2017; 63(6): 403-408
Category:
Původní práce
Summary
Objective:
The aim of proposed paper was to compare a three total 25-hydroxy-vitamin D immunoassays to that of HPLC with UV detection.
Material and methods:
Serum 25-(OH) D levels were measured from blood samples of 109 patients with different immunoassays (ABBOTT, ROCHE, SIEMENS) and method of HPLC which was chosen as the reference. In the first step immunoassays were compared to HPLC. In the second step immunoassays were compared to each other. Further purpose of methods comparison the Passing-Bablok regression and Bland-Altman analysis were used. The limits of maximum acceptable differences were set at 21.5 %, according to Vitamin D Standardization-Certification Program (VDSCP). In the last step, the concordance in the interpretation of measured results was evaluated.
Results:
None of the examined 25-(OH) D immunoassays was comparable to HPLC and to each other. Bland-Altman analysis revealed, in comparison to HPLC, that ROCHE showed positive bias +28.0 %, ABBOTT +0.2 % and SIEMENS –23.4 %. Although average bias of ABBOTT immunoassay is insignificant, particular results do deviate significantly (-89.4 % to 89.0 %). The concordance in the interpretation of measured results, in comparison to HPLC, was highest with ABBOTT (65.21 %), then with ROCHE (59.63 %) and lowest with SIEMENS (47.79 %).
Conclusion:
The results of the proposed papers suggest low levels of 25-(OH) D immunoassays standardization and an alternative to use assay-specific decision limits.
Keywords:
25-hydroxy-vitamin D – interpretation – method comparison – standardization
Zdroje
1. Cannell JJ, Hollis BW, Zasloff M et al. Diagnosis and treatment of vitamin D deficiency. Expert Opin Pharmacother 2008; 9(1): 107–118.
2. Huibers MH, Visser DH, Deckers MM et al. Vitamin D deficiency among native Dutch and first – and second generationnon – Westernimmigrants. Eur J Pediatr 2014; 173(5): 583–588. Dostupné z DOI: <http://dx.doi.org/10.1007/s00431–013–2198-x>.
3. Bischoff-Ferrari HA, Dietrich T, Orav EJ et al.l Positive association between 25-hydroxy vitamin D levels and bone mineral density: a population-based study of younger and older adults. Am J Med 2004; 116(9): 634–639.
4. Murad MH, Elamin KB, Abu Elnour NO et al. Clinical review: The effect of vitamin D on falls: a systematic review and meta-analysis. J Clin Endocrinol Metab 2011; 96: 2997–3006. Dostupné z DOI: <http://dx.doi.org/10.1210/jc.2011–1193>.
5. Bentley J. The role of vitamin D in infants, children and young people. Nurs Child Young People 2015; 27(1): 28–35; quiz 36. Dostupné z DOI: <http://dx.doi.org/10.7748/ncyp.27.1.28.e508>.
6. Pittas AG, Lau J, Hu FB et al. The role of vitamin D and calcium in type 2 diabetes. A systematic review and meta-analysis. J Clin Endocrinol Metab 2007; 92(6): 2017–2029.
7. Forman JP, Bischoff-Ferrari HA, Willett WC et al. Vitamin D intake and risk of incident hypertension: results from three large prospective cohort studies. Hypertension 2005; 46: 676–682.
8. Grant WB. How strong is the evidence that solar ultraviolet B and vitamin D reduce the risk of cancer? An examination using Hill’s criteria for causality. Dermatoendocrinol 2009; 1(1): 17–24.
9. Melamed ML, Michos ED, Post W et al. 25-hydroxyvitamin D levels and the risk of mortality in the general population. Arch Intern Med 2008; 168(15): 1629–1637. Dostupné z DOI: <http://dx.doi.org/10.1001/archinte.168.15.1629>.
10. Garland CF, Gorham ED, Mohr SB et al. Vitamin D for cancer prevention: global perspective. Ann Epidemiol 2009; 19(7): 468–483. Dostupné z DOI: <http://dx.doi.org/10.1016/j.annepidem.2009.03.021>.
11. Saenger AK, Laha TJ, Bremner DE et al. Quantification of serum 25-hydroxyvitamin D(2) and D(3) using HPLC-tandem mass spectrometry and examination of reference intervals of or diagnosis of vitamin D deficiency. Am J Clin Pathol 2006; 125(6): 914–920.
12. Ross AC, Taylor CL, Yaktine AL et al (eds). [Institute of Medicine (IOM)]. Dietary Reference Intakes for Calcium and Vitamin D. The National Academic Press: Washington (DC) 2011. ISBN 978–0-309–16394–1.
13. Seamans KM, Cashman KD. Existing and potentially novel functional Markers of vitamin D status: a systematic review. Am J Clin Nutr 2009; 89(6): 1997S-2008S. Dostupné z DOI: <http://dx.doi.org/10.3945/ajcn.2009.27230D>.
14. Kwak HS, Chung HJ, Cho DH et al. Efficacy of the measurement of 25-hydroxyvitamin D2 and D3 levels by using Perkin Elmer liquid chromatography-tandem mass spectrometry vitamin D kit compared with Dia Sorinradio immunoassay kit and Elecsys vitamin D total assay. Ann Lab Med 2015; 35(2): 263–265. Dostupné z DOI: <http://dx.doi.org/10.3343/alm.2015.35.2.263>.
15. Tsugawa N, Suhara Y, Kamao M et al. Determination of 25-hydroxyvitamin D in human plasma using high-performance liquid chromatography – tandem mass spectrometry. Anal Chem 2005; 77(9): 3001–3007.
16. Wallace AM, Gibson S, de la Hunty A et al. Measurement of 25-hydroxyvitamin D in the clinical laboratory: current procedures, performance characteristics and limitations. Steroids 2010; 75(7): 477–488. Dostupné z DOI: <http://dx.doi.org/10.1016/j.steroids.2010.02.012>.
17. Janssen MJ, Wielders JP, Bekker CC et al. Multicenter comparison study of current methods to measure 25-hydroxyvitamin D in serum. Steroids 2012; 77(13): 1366–1372. Dostupné z DOI: <http://dx.doi.org/10.1016/j.steroids.2012.07.013>.
18. Carter GD. 25-Hydroxyvitamin D assays: the quest for accuracy. Clin Chem 2009; 55(7): 1300–1302. Dostupné z DOI: <http://dx.doi.org/10.1373/clinchem.2009.125906>.
19. Binkley N, Krueger D, Cowgill CS et al. Assay variation confounds the diagnosis of hypovitaminosis D: a call for standardization. J Clin Endocrinol Metab 2005; 89(7): 3152–3157.
20. Binkley N, Krueger D, Gemar D et al. Correlation among 25-hydroxy-vitamin D assays. J Clin Endocrinol Metab 2008; 93(5): 1804–1808. Dostupné z DOI: <http://dx.doi.org/10.1210/jc.2007–2340>.
21. Siemens Advia Centaur® XP. Celkový vitamín (Vit D) [direction. insert 0699279_SK Rev D]. Siemens Healthcare Diagnostics: Tarrytown (NY) 2016–04.
22. Architect® 25-OH Vitamin D [directional insert 3L52, G2–4021/R03]. Abbott Laboratories. Wiesbaden 2012. Dostupné z WWW: <http://www.ilexmedical.com/files/download/1306390706e06Sh.pdf>.
23. Roche ELECSYS®. Vitamin D total [directional insert 05894913 190, V 7.0 Slovenskz]. Roche Diagnostics GmbH: Mannheim 2016–08.
24. Immunodiagnostik 25-OH-Vitamin D3/D2 RP-HPLC Kit [directional insert KC3000], Immundiagnostik AG: Bensheim 2015–06. Dostupné z WWW: http://www.immundiagnostik.com/fileadmin/pdf/KC3000%20Vitamin%20D3%20D2_RP.pdf
25. Pludowski P, Karczmarewicz E, Bayer M et al. Practical guidelines for the supplementation of vitamin D and the treatment of deficits in Central Europe – recommended vitamin D intakes in the general population and group sat risk of vitamin D deficiency. Endokrynol Pol 2013; 64(4): 319–327.
26. Holick MF, Binkley NC, Bischoff-Ferrari HA et al. Endocrine Society. Evaluation, treatment, and prevention of vitamin D deficiency: an Endocrine Society clinical practice guideline. J Clin Endocrinol Metab 2011; 96(7): 1911–1930. Dostupné z DOI: <http://dx.doi.org/10.1210/jc.2011–0385>. Erratum in J Clin Endocrinol Metab 2011; 96(12): 3908.
27. Ross AC, Taylor CL, Yaktine AL et al. (eds). [Institute of Medicine (US) Committee to Review Dietary Reference Intakes for Vitamin D and Calcium]. Dietary Reference Intakes for Calcium and Vitamin D. National Academies Press: Washington (DC): 2011. ISBN 978–0-309–16394–1.
28. Cavalier E, Dozet E, Gadisseur R et al. Measurement uncertainty of 25-OH vitamin D determination with different commercially available kits: impact on the clinical cutoffs. Osteoporosis Int 2010; 21(6): 1047–1051. Dostupné z DOI: <http://dx.doi.org/10.1007/s00198–009–1052–5>.
29. Cavalier E, Carlisi A, Beckert A et al. Analytical evaluation of the new Abbott Architect 25-OH vitamin D assay. Clin Biochem 2012; 45(6): 505–508. Dostupné z DOI: <http://dx.doi.org/10.1016/j.clinbiochem.2012.01.021>.
30. Lei JK, Lucas RM, Clements MS et al. Assessing vitamin D status: pit falls for the un way. Mol Nutr Food Res 2010; 54(8): 1062–1071. Dostupné z DOI: <http://dx.doi.org/10.1002/mnfr.200900468>.
31. Roth HJ, Schmidt-Gayk H, Weber H et al. Accuracy and clinical implications of seven 25-hydroxyvitamin D compared with liquid chromatography – tandem mass spectrometry as reference. Ann Clin Biochem 2008; 45(Pt 2): 153–159. Dostupné z DOI: <http://dx.doi.org/10.1258/acb.2007.007091>.
32. Farrell CJ L, Martin S, Mc Whinney B et al. State of Art vitamin D assays: A comparison of Automated Immunoassays with Liquid chromatography-Tandem Mass Spectrometry Methods. Clin Chem 2012; 58(3): 531–542. Dostupné z DOI: <http://dx.doi.org/10.1373/clinchem.2011.172155>.
33. Ong L, Saw S, Schabdeen NB et al. Current 25-hydroxyvitamin D assays: Do the ypass the test? Clin Chim Acta 2012; 413(13–14): 1127–1134. Dostupné z DOI: <http://dx.doi.org/10.1016/j.cca.2012.03.009>.
34. Snellman G, Melkus H, Godeborg R et al. Determining vitamin D status, a comparison between commercially available assays. Plos One 2010; 5(7): e11555. Dostupné z DOI: <http://dx.doi.org/10.1371/journal.pone.0011555>.
35. Friedecký B, Vávrová J. Současný stav stanovení vitaminu D v séru. Klin Biochem Metab 2012; 20[41](3): 174–178 .Dostupné z WWW: <http://www.cskb.cz/res/file/KBM-pdf/2012/2012–3/KBM12–3-Fr-vitD-174.pdf>.
36. Carter GD. 25-hydroxyvitamin D: A difficult analyte. Clin Chem 2012; 58(3): 486–488. Dostupné z DOI: <http://dx.doi.org/10.1373/clinchem.2011.180562>.
37. Kricka LJ. Human anti-animal antibody interferences in immunological assays. Clin Chem 1999; 45(7): 942–956.
Štítky
Diabetologie Endokrinologie Interní lékařstvíČlánek vyšel v časopise
Vnitřní lékařství
2017 Číslo 6
- Není statin jako statin aneb praktický přehled rozdílů jednotlivých molekul
- Testování hladin NT-proBNP v časné diagnostice srdečního selhání – guidelines ESC
- Pregabalin je účinné léčivo s příznivým bezpečnostním profilem pro pacienty s neuropatickou bolestí
- Syndrom Noonanové: etiologie, diagnostika a terapie
- Moje zkušenosti s Magnosolvem podávaným pacientům jako profylaxe migrény a u pacientů s diagnostikovanou spazmofilní tetanií i při normomagnezémii - MUDr. Dana Pecharová, neurolog
Nejčtenější v tomto čísle
- Primární testikulární lymfomy
- Úskalia stanovenia vitamínu D
- Inzulinová pumpa u diabetu 2. typu: léčba zaměřená na B-buňku
- Trombóza a malignita – editorial