The Standard of Analytical Quality of Glucometers
Authors:
B. Friedecký 1,2; D. Springer 3; J. Kratochvíla 2
Authors‘ workplace:
ÚKBD FN Hradec Králové
1; SEKK s. r. o. Pardubice
2; ÚLBLD FVL Praha
3
Published in:
Klin. Biochem. Metab., 22 (43), 2014, No. 2, p. 74-80
Overview
Review on the recent state in measurement of blood glucose by glucometers. Summarization of requirements on the precision and bias of results. Describing of the new quality standard ISO 15197:3013. Introduction of some results in glucometers testing in some countries and publications. Main emphasis is given to self-monitoring of patients more briefly is introduced glucometers using in the ambulances, in the intensive care in hospitals and continual glucose monitoring. The aim of publication is to help in improvement quality diabetes care and economy of use for the management of diabetes and intensive care.
Keywords:
glucometers, ISO 15197:2013, quality, self monitoring, diabetes.
Sources
1. POCT 12 A3. Point-Of-Care Blood Glucose Testing in Acute and Chronic Care Facilities. Approved Guideline-Third Edition. Clinical Laboratoty Standard Institute 2013.
2. Schnell, O., Erbach, M., Wintergerst, E. Higher accuracy of self-monitoring of blood glucose in insulin treated patients in Germany: clinical and economical aspects. J Diabetes Sci. Technol. 2013, 7, p. 904-912.
3. EN ISO 15197:2013. In vitro diagnostic test systems – Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. Geneve 2013.
4. Bernstein, R., Parkes, G. L., Goldy, A., Brown, D., Harrison, B. at al. A new test strip technology platform for self-monitoring of blood glucose. J Diabetes Sci. Technol. 2013, 7, p. 1386-1399.
5. Freckmann, G., Baumstark, A., Jendrike, N., Zschornack, E., Kocher, S. et al. System accuracy evaluation of 27 blood glucose monitoring systems according to DIN EN ISO 15197. Diabetes Technol. Ther. 2010, 12(3), p. 221-231.
6. Freckmann, G., Schmid, C., Baumstark, A., Pleus, S., Link, M. et al. System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197. J Diabetes Sci. Technol. 2012, 1(6), p. 1060-1075.
7. Freckmann, G., Baumstark, A., Schmid, C., Pleus, S., Link, M., Haug, C. Evaluation of 12 blood glucose monitoring systems for self-testing: system accuracy and measurement repeatability. Diabetes Technol. Ther. 2014, 16(2), p. 113-122.
8. Garg, S. K., Hirsch, I. B. Self-monitoring of Blood Glucose-An Overwiew. Diabetes Technol. Ther. 2014, 16, Suppl,1, E pub.
9. Baumstark, A., Pleus, S., Schmid, C., Link, M., Haug, C. Lot-to lot variability of test strips and accuracy assessment of systems for self-monitoring for blood glucose according to ISO 15197. J Diabetes Sci. Technol. 2012, 6/5, p. 1076-1086.
10. Brazg, R. L., Kleff, L. D., Partin, C. C. Performance variability of serum commonly used for self-monitoring of blood glucose systems:clinical considerations for patients and providers.J Diabetes Sci. Technol. 2013, 7, p. 144-152
11. Ramljak, S., Lick, J. P., Schipper, C. et al. Hematocrit interference of blood glucose meters for patient self-monitoring. J Diabetes Sci. Technol. 2013, 7, p. 179-189.
12. Koschinsky, T. et al. Sind POCT-Systeme für die Dia-gnostik des Gestationsdiabetes geeignet? Kongress DGKL Dresden 23 – 26. 10 2013. Clin. Chem. Lab. Med. 2013, 51(10), eA1 – eA85.
13. Sacks, D. B., Arnold, M., Baktris, G. L., Bruns, D. E., Horvath, A.R. et al. Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Clin. Chem. 2011, 57, p. 793-798.
14. Friedecký, B., Kratochvíla, J. SKUP – cesta ke zvýšení kvality péče o pacienty. Prakt. Lék. 2012, 92(10), p. 492-495.
15. http//www.skup.nu. Dostupné on line 12.3.2014.
16. Krinsley, J. S. Understanding glycemic control in the critically ill: 2011 update. Hosp. Pract. 2011, 39(2), p. 47-55.
17. O´Kane, M. J. The accuracy of point-of-care glucose measurement. Ann. Clin. Biochem. 2012, 49, p. 108-109.
18. Krinsley, J. S., Egi, M., Kiss, A., Devedra, A. N., Schuetz, P., Maurer, P. M. et al. Diabetic status and the relation of three domains of glycemic control to mortality in critically ill patients: an international multi-center cohort study. Crit. Care 2013, 1, 17(2), R37.
19. Karon, B. S., Boyd, J. C., Klee, G. G. Glucose meter performance criteria for tight glycemic control estimated by simulation modelling. Clin. Chem. 2010, 56(7), p. 1091-1097.
20. Scott, M. G., Bruns, D. E., Boyd, J. C., Sacks, D. B. Tight glucose control in the intensive care unit. Are glucose meters up to the task? Clin. Chem. 2009, 55, p. 18-20.
21. Gielen, M., Vlaselaers, D., Van den Berghe, G. Glucose in the ICU-evidence, guideliness and outcomes. N. Engl. J. Med. 2012, 20, p. 2451-2453.
22. Boyd, J. S., Bruns, D. E. Effects of measurement frequency on analytical quality required for glucose measurements in intensive care units“assessment by simulation models. Clin. Chem. 2014, 60, E pub
23. Van den Berghe, G. Blood glucose monitoring in the intensive care unit:toward defining and imprecision thresholds (near) continuous sensors. Clin. Chem. 2014, 60, E pub.
24. Heinemann, L., DeVries, J. H. Evidence for continuous glucose monitoring: sufficient for reimbuinsment? Diabet. Med. 2014, 3, p. 122-125.
25. Barthelino, T., Bode, B. W. Continuous Glucose Monitoring in 2013. Diabetes Technol. Ther. 2014, 1(16), E pub.
26. Leelarathma, L., English, S. W., Thabit, H., Caldwell, K., Allen, J. M. et al. Accuracy of Subcutaneous Continuous Glucose Monitoring in Critically Ill Adults. Improved Sensor Performance with Enhance Calibrations. Diabetes Technol. Ther. 2014, 1, p. 93-101.
27. Freckmann, G., Pleus, S., Link, M., Zschornack, E., Klotzer, H. M., Haug, C. Performance evaluation of three continuous glucose monitoring systems:comparison of six sensors per subject in parallel. J Diabetes Sci. Technol. 2013, 7, p. 842-852.
Labels
Clinical biochemistry Nuclear medicine Nutritive therapistArticle was published in
Clinical Biochemistry and Metabolism
2014 Issue 2
Most read in this issue
- Clinical Problems of IgM Monoclonal Gammapathies
- Risk factors of Asymptomatic (Smoldering) Multiple Myeloma
- Diagnosis and stratification of systemic AL amyloidosis in the light of the 2013 Czech Myeloma Group recommendation
- The Standard of Analytical Quality of Glucometers