SCREW LOOSENING IN ABUTMENT: CAUSES, ORIGIN MECHANISM, COMPLICATIONS AND MANAGEMENT


Authors: J. Staněk 1;  K. Singh 2;  A. Jusku 1;  M. Hammal 1;  N. Králová 1;  B. Hocková 3;  B. Azar 1
Authors‘ workplace: Klinika zubního lékařství, Lékařská fakulta Univerzity Palackého a Fakultní nemocnice, Olomouc 1;  Bhandal Dental Practice, Cradley Heath, Anglie 2;  Oddelenie maxilofaciálnej chirurgie, Fakultná nemocnica s poliklinikou F. D. Roosevelta, Banská Bystrica 3
Published in: Česká stomatologie / Praktické zubní lékařství, ročník 120, 2020, 4, s. 116-121
Category: Review Article

Overview

Introduction and goal: Tooth replacement with dental implants is an invasive method, demanding in time and money. For these reasons, it is important to understand the technical complications that occurs in five years in about 12% of crowns on dental implants. Loosening of the screw in the abutment can have serious consequences for the patient, from the destruction of the crown when trying to find the fixation screw, associated with bacterial colonization leading to perimplantitis or its accentuation, to the fracture of the screw or fixture.

The aim of the presented communication is to present current knowledge about the causes of screw loosening in the abutment, their prevention and a suitable method of treatment.

Conclusion: Loosening the screw in the abutment is a common occurrence with adverse consequences that may jeopardize functionality, aesthetics and treatment success. This is not a devastating complication, but recurrence is not desirable.

A loose screw can be an early warning of inadequate replacement biomechanics, or implant overload. Accurate diagnosis is also necessary so that an implant with loss of osseointegration is not considered a loose crown on the implant and vice versa.

Keywords:

loosening of the screw in the abutment – settling effect – complications


Sources

1. Kourtis S, Kaitatzidou S, Kaitatzidou A, Roussou V. Loosening of the fixing screw in single implant crowns: predisposing factors, prevention and treatment options. J Esthet Restor Dent. 2017; 29(4): 233–246.

2. Jung RE, Zembic A, Pjetursson BE, et al. Systematic review of the survival rate and the incidence of biological, technical, and aesthetic complications of single crowns on implants reported in longitudinal studies with a mean follow-up of 5 years. Clin Oral Implants Res. 2012; 23: 2–21.

3. Zembic A, Kim S, Zwahlen M, Kelly JR. Systematic review of the survival rate and incidence of biologic, technical, and esthetic complications of single implant abutments supporting fixed prostheses. Int J Oral Maxillofac Implants. 2014; 29: 99–116.

4. Goodacre CJ, Kan JY, Rungcharassaeng K.Clinical complications of osseointegrated implants. J Prosthet Dent. 1999; 81: 537–552.

5. Katsavochristou A, Koumoulis D. Incidence of abutment screw failure of single or splinted implant prostheses: A review and update on current clinical status. J Oral Rehabil. 2019; 46(8): 776–786.

6. Jung RE, Pjetursson BE, Glauser R, Zembic A, Zwahlen M, Lang NP. A systematic review of the 5-year survival and complication rates of implant-supported single crowns. Clin Oral Implants Res. 2008; 19(2): 119–130. Epub 2007 Dec 7.

7. Gracis S, Michalakis K, Vigolo P, et al.Internal vs. external connections for abutments/reconstructions: a systematic review. Clin Oral Implants Res. 2012; 23: 202–216.

8. Gaddale R, Mishra SK, Chowdhary R. Complications of screw – and cement-retained implant-supported full-arch restorations: A systematic review and meta-analysis. Int J Oral Implantol. 2020; 13(1): 11–40.

9. Jatin KJ, Sethuraman R, Chauhan S, Javiya P, Srivastava S, Patel R, Bhalani B.Retention failures in cement- and screw-retained fixed restorations on dental implants in partially edentulous arches: A systematic review with meta-analysis J Indian Prosthodont Soc. 2018; 18(3): 201–211.

10. Wang JH, Judge R, Bailey D. A 5-year retrospective assay of implant treatments and complications in private practice: the restorative complications of single and short-span implant-supported fixed prostheses. Int J Prosthodont. 2016; 29(5): 435–444.

11. Kourtis S. Long term clinical results from the use of implants. A clinical research monography. Athens, 2010.

12. Korsch M, Walther W. Retrospective analysis of loosening of cement-retained vs screw-retained fixed implant-supported reconstructions. Quintessence Int. 2015; 46(7): 583–589.

13. Lang LA, May KB, Wang RF. The effect of the use of a countertorque device on the abutment-implant complex. J Prosthet Dent. 1999; 81: 411–417.

14. Resnik RR. Fixed prosthodontics complications. In: Misch CE, Misch Avoiding complications in oral implantology. 1. vydání. St. Luise: Mosby Elsevier; 2017, 631–634.

15. Binon PP. The effect of implant/abutment hexagonal misfit on screw joint stability. Int J Prosthodont. 1996; 9: 149–160.

16. Binon PP. Evaluation of three slip fit hexagonal implant. Implant Dent, 1996; 5: 235–248.

17. Boggan S, Strong JT, Misch CE, et al. Influence of hex geometry and prosthetic table width on static and fatique strenth of dental implants. J Prosthet Dent 1999; 82: 436–440.

18. Cho SC, Small PN, Elian N, Tarnow D.Screw loosening for standard and wide diameter implants in partially edentulous cases: 3- to 7-years longitudinal data. Implant Dent. 2004; 13: 245–250.

19. Siamos G, Winkler S, Boberick KG. Relationship between implant preload and screw loosening on implant-supported prostheses. J Oral Implantol. 2002; 28(2): 67–73.

20. Shadid R, Sadaqa N. A comparison between screw- and cement-retained implant prostheses. a literature review. J Oral Implantol. 2012; 38(3): 298–307.

21. Theoharidou A, Petridis HP, Tzannas K, Garefis P. Abutment screw loosening in single-implant restorations: a systematic review. Int J Oral Maxillofac Implant. 2008; 23: 681–690.

22. Weber HP, Sukotjo C. Does the type of implant prosthesis affect outcomes in the partially edentulous patient? Int J Oral Maxillofac Implants. 2007; 22: 140–172.

23. Freitas-Junior AC, Almeida EO, Bonfante EA, et al. Reliability and-failure modes of internal conical dental implant connections. Clin Oral Implants Res. 2012; 24: 197–202.

24. Schmitt CM, Nogueira-Filho G, Tenenbaum HC, et al. Performance of conical abutment (Morse Taper) connection implants: A systematic review. J Biomed Mater Res Part A. 2014; 102A: 552–574.

25. McGlumphy EA, Mendel DA, Holloway JA.Implant screw mechanics. Dent Clin N Am. 1998; 42: 71–89.

26. Winkler S, Ring K, Ring JD, Boberick KG. Implant screw mechanics and the settling effect: overview. J Oral Implantol. 2003; 29(5): 242–245.

27. Oberg E, Franklin DJ, McCauley CJ Heald, Ricardo M. Machinery's Handbook, 27th  ed. Industrial Press, 2004.

28. Dellinges MA, Tebrock OC. A measurement of torque values obtained with hand-held drivers in a simulated clinical setting. Int J Prosthodont. 1993; 2: 212–214.

29. Sakaguchi RL, Borgersen SE. Nonlinear contact analysis of preload in dental implant screws. Int J Oral Maxillofac Implants. 1995; 10: 295–302.

30. Jorneus L, Jemt T, Carlsson L. Loads and designs of screw joints for single crowns supported by osseoin. Sheldon Winkler et al. Journal of Oral Implantology: 245 tegrated implants. Int J Oral Maxillofac Implants. 1992; 7: 353–359.

31. Resnik RR. Fixed prosthodontics complications. In: Misch CE, Mischs Avoiding complications in oral implantology. 1. vydání. St. Luise: Mosby Elsevier; 2017, 631–645.

32. Nigro F, Sendyk CL, Francischone CE Jr, Francischone CE. Removal torque of zirconia abutment screws under dry and wet conditions. Braz Dent J. 2010; 21(3): 225–228.

33. Yao KT, Kao HC, Cheng CK, et al. The effect of clockwise and counterclockwise twisting moments on abutment screw loosening. Clin Oral Implants Res. 2011; 23: 1–6.

34. Weiss EI, Kozak D, Gross MD, Goldschleger G, Goldschleger M. Effect of repeated closures on opening torque values in seven abutment-implant systems. J Prosthet Dent. 200; 84(2): 194-9. doi: 10.1067/mpr.2000.108069.

35. Hebel KS, Gajjar RC. Cement-maintained versus restoration of the implant by means of a screw: achieving optimal occlusion and aesthetics in implant dentistry. J Prosthet Dent. 1997; 77: 28–35.

36. Balshi TJ, Hernandez RE, Pryszlak MC, et al.A comparative study of one implant versus two replacing a single molar. Int J Oral Maxillofac Implants. 1996; 11: 372–378.

37. Yilmaz B, Seidt JD, McGlumphy ED, Clelland NL. Comparison of strains for splinted and nonsplinted screw-retained prostheses on short implants. Int J Oral Maxillofac Implants. 2011; 26: 1176–1182.

38. Misch CE, Bidez MW. Implant protected occlusion: a biomechanical rationale. Compend Contin Dent Educ. 1994; 15: 1330–1343.

39. Hurson S. Practical clinical guidelines to prevent screw loosening. Int J Dent Symp. 1995; 3(1): 23–25.

40. Schoenbaum T, Wadhwani C, Stevenson RG.Covering the implant prosthesis screw acces hole: A biological approach to material selection and technique. J Oral Implant. 2017; 43(1): 39–44.

41. Canullo L, Penarrocha-Oltra D, Covani U. Microbiological assessment of the implant-abutment interface in different connections: cross-sectional study after 5-years of functional loading. Clin Oral Implants Res. 2015; 26: 426–434.

42. Wadhwani C, Pineyro A, Avots J. An esthetic solution to the screw retained implant restoration: introduction to the implant crown adhesive plug: clinical report. J Esthet Restor Dent. 2011; 23: 138–143.

43. Goodacre CJ, Kan JYK, Rungcharassaeng K. Clinical complications of osseointegrated implants. J Prosthet Dent. 1999; 81: 537–552.

Labels
Maxillofacial surgery Orthodontics Dental medicine
Topics Journals
Login
Forgotten password

Enter the email address that you registered with. We will send you instructions on how to set a new password.

Login

Don‘t have an account?  Create new account

#ADS_BOTTOM_SCRIPTS#