Verification of the reference range of free triiodothyronine in serum on the analyzer Architect ci 16200


Authors: T. Vacková;  J. Franeková;  A. Jabor
Authors‘ workplace: Praha ;  Oddělení klinické biochemie, Pracoviště laboratorních metod, Institut klinické a experimentální medicíny
Published in: Klin. Biochem. Metab., 26, 2018, No. 4, p. 178-181

Overview

Objective:

Verification of the reference interval of anew method for the determination of free triiodothyronine in serum on the Abbott Architect 16200 analyzer.

Design:

Original Article.

Settings:

Institute for Clinical and Experimental Medicine, Vídeňská 1958/9, 140 21 Praha

Material and methods:

The reference interval was validated on atotal of 110 serum samples from healthy subjects, comprising 55 females and 55 males. All samples used were from the IKEM Biobanka. The FT3_6 microparticle chemiluminescent immunoassay Abbott kit was used for the assay.

Results:

All free serum triiodothyronine values measured by the new fT3_6 method were in the reference interval indicated by the manufacturer. The reference range determined by anon-parametric statistical method with the Box-Cox transformation on the IKEM Biobanka samples was 3.21 - 4.78 pmol/L.

Conclusion:

The fT3 reference range established by the Abbott’s new fT3_6 method can be used in clinical practice with the range specified by the manufacturer.

Keywords:

Architect 16200, fT3, reference interval (RI).


Sources

1. Abbott Diagnostic Division Informace oproduktu ARCHITECT Free T3, Ireland, 13. 7. 2017, 11 s.

2. Abbott Diagnostic Division Příbalový leták kproduktu ARCHITECT Free T3 platný pro katalogová čísla 7K63-27(37, 32), česká verze G8-1469/R03, 03/2017, 6 s.

3. Límanová, Z., Pikner, R., Springer, D. Doporučení pro laboratorní diagnostiku funkčních aautoimunitních onemocnění štítné žlázy. Klin. Biochem. Metab., 19 (40), 2011, No. 1, p. 48-61.

4. Demers, L.M., Spencer, C.A. Laboratory Support for the Diagnosis and Monitoring of Thyroid Disease 2002 NACB, Laboratory Medicine Practise Guidelines, 2002, 13, p. 125

5. Zima, T. Laboratorní diagnostika. Praha: Galén, 2013, 1146 s., ISBN 9788074920622

6. Thomas, L. Clinical Laboratory Diagnostics: Use and Assessment of Clinical Laboratory Results. Frankfurt am Main: TH-Books, 1998, 1527 s. ISBN 3-9805215-4-0.

7. Röcker, L., Janiszewski, Ch., Hopfenmüller, W. Determination of reference ranges for thyroid hormones in serum for the population of Berlin (West) with special emphasis of age- and gender- dependence. Lab. Med. 1988, 12, p. 213-22

8. Liappis, N., Bernstorff v, W., Schlebusch, H. Refe-rence values for the concentration of free triiodothyronine, triiodothyronine, free thyroxine, thyroxine, thyrotropin and thyroxin-binding globulin in serum of euthyroid adults. Influence of oral contraceptives. Method: luminiscence-enhanced enzyme immunoassay. Ärztl. Lab. 1990, 36, p. 8-11

9. Racek, J. Klinická biochemie. Praha: Galén, 2006, 329 s. ISBN 8072623249

10. Abbott Diagnostic Division Příbalový leták kproduktu ARCHITECT Free T4 platný pro katalogová čísla 7K65-39 (24, 39, 34), česká verze G8-1500/R03, 04/2017, 6 s.

11. Abbott Diagnostic Division Příbalový leták kproduktu ARCHITECT TSH platný pro katalogová čísla 7K62-35 (20, 25, 30), česká verze G6-0153/R06, 11/2015, 7 s.

Labels
Clinical biochemistry Nuclear medicine Nutritive therapist
Topics Journals
Login
Forgotten password

Enter the email address that you registered with. We will send you instructions on how to set a new password.

Login

Don‘t have an account?  Create new account

#ADS_BOTTOM_SCRIPTS#