Technical note on quality control tests of dose calibrators
Authors:
I. Šimperský 1,2; D. Pecinová 1
Authors‘ workplace:
Oddělení nukleární medicíny, FN Hradec Králové
1; Klinika radiologie a nukleární medicíny, FN Brno – Bohunice, ČR
2
Published in:
NuklMed 2023;12:42-45
Category:
Original Article
Overview
Quality control tests of dose calibrators are an integral part of routine practice in all departments of nuclear medicine. They should be carried out regularly to ensure their optimum performance at all times and to determine the rate and extend of any deterioration in time. However, when there is no additional software purchased for these particular purposes, data collecting, and subsequent processing could sometimes be a tedious and time-consuming process. Based on the requirements of clinical practice an efficient method was developed to facilitate data handling. The described technical solution enables direct reading of data from the dose calibrator to a computer (tablet, smartphone, etc.) and their subsequent immediate processing. A conversion unit using Bluetooth technology was designed for communication between the measuring device and the computer. Communication with the ionization chamber is controlled by a macro written in VBA (Visual basic for applications). The data loaded into Excel are subsequently processed and evaluated according to the user's requirements. The presented solution is cheap and easy to modify, its big advantage is the compatibility with MS Office, which is currently easily accessible to all users at nuclear medicine workplaces.
Keywords:
Curiementor – dose calibrator – Bluetooth – test of background response – test of precision – test of accuracy – Visual basic for aplications – VBA
Sources
- Vyhláška č. 422/2016 Sb., o radiační ochraně a zabezpečení radionuklidového zdroje. In: Sbírka zákonů České republiky. 2016, částka 172.
- SÚJB. Radiační ochrana v nukleární medicíně – systém kontrol detekční a zobrazovací techniky [Online]. 2019. [cit. 2023-03-31]. Dostupné na: https://www.sujb.cz.
- PTWFreiburg. User Manual CURIEMENTOR 3/4 RS232 Interface, D704.131.01/02, 2006-10.
- Microsoft. Identifying Functions in DLLs (kernell.dll) [Online]. 2022. [cit. 2023-03-31]. Dostupné na: https://learn.microsoft.com/en-us/dotnet/framework/interop/identifying-functions-in-dlls.
Labels
Nuclear medicine Radiodiagnostics RadiotherapyArticle was published in
Nuclear Medicine
2023 Issue 3
Most read in this issue
- Technical note on quality control tests of dose calibrators
- Effect of lymphodrainage of lower extremities in patients with a diabetic foot syndrome – a pilot study
- AtomLab 500 Radionuclide Calibrator: Advantages and Disadvantages of Built-in Attenuator Tube Linearity Test vs. Source Decay Method
- Nukleární medicína a MUDr. Eva Hoffmannová, CSc Kolín