Results of the Cataract Surgeries with the Acrysoft Restor SN6AD3 Intraocular Lens Implantation
Authors:
K. Marešová; P. Mlčák; O. Vláčil
Authors‘ workplace:
Oční klinika LF UP a FN, Olomouc, přednosta doc. MUDr. Jiří Řehák, CSc.
Published in:
Čes. a slov. Oftal., 66, 2010, No. 1, p. 26-28
Overview
Purpose:
this study aims to evaluate the postoperative results of uncorrected visual acutity (UCVA) near and distance and contrast sensitivity after cataract surgery with Acrysof ReSTOR SN6AD3 implantation and to evaluate the personal questionnaire.
Methods:
this prospective study included 40 eyes of 21 patients with cataract. Each patient had small incision cataract surgery with Acrysof ReSTOR SN6AD3 implantation. Corrected and uncorrected distance visual acuity, corrected and uncorrected close visual acuity and corrected scotopic contrast sensitivity were measured before and after surgery. The subjective impressions of patients were evaluated by a personal questionnaire.
Results:
mean distance UCVA before surgery was 0.3 ± 0.19, mean distance UCVA after surgery was 0.9 ± 0.3. Mean near UCVA before surgery was 11.9 ± 2.2, mean near UCVA after surgery was 1.8 ± 1.1. Contrast sensitivity before surgery was 25.6 ± 10.3 and after surgery was 33.3 ± 1.8. The evaluation of the questionnaire revealed that 19 patients (95%) never wear glasses, 17 patients (80%) are fully satisfied with reading very small text and 17 patients (80%) are fully satisfied with far vision. Hallos around spot lights recorded 12 patients (60%). Problems with contrast noticed 3 patients (15%). All of the patients would prefer the implantation of the same intraocular lens and recommend it to friend.
Conclusion:
The results shows that uncorrected both distance and near visual acuity were better after surgery (p < 0.0001). Contrast sensitivity also improved after surgery (p = 0.0001). Personal questionnaire revealed high satisfaction after surgery.
Key words:
cataract surgery, multifocal intraocular lenses, Acrysof ReSTOR.
Sources
1. Blaylock, J.,F., Si, Z., Vickers, Ch.: Visual and refractive status at different focal distances after implantation of the ReSTOR multifocal intraocular lens. J Cataract Refract Surg, 2006; 32: 1464–1473.
2. Cionni, J.,R., Osher, R.,H., Snyder, M.,E., et al.: Visual outcome comparison of unilateral versus bilateral implantation of apodized diffractive intraocular lens implantation after cataract extraction: Prospective 6 month study. J Cataract Refract Surg, 2009; 35: 1033–1039.
3. Cuq, C., Spera, C., Laurendeau, C., et al.: Intermediate visual acuity without spectacles following bilateral ReSTOR implantation. Eur J Ophthalmol, 2008; 18: 733–738.
4. Ferrer-Blasco, T., Montés-Micó, R., CerviĖo, A.: Contrast sensitivity after refractive lens exchange with diffractive multifocal intraocular lens implantation in hyperopic eyes. J Cataract Refract Surg, 2008; 34: 2043–2048.
5. Chang, D.,F.: Prospective functional and clinical comparison of bilateral ReZoom and ReSTOR intraocular lenses in patiens 70 years or younger. J Cataract Refract Surg, 2008; 34: 934–941.
6. Chiam, J.,T., Chan, J.,H., Aggarwal, R.,K., et al.: ReSTOR intraocular lens implantation in cataract surgery: Quality of vision. J Cataract Refract Surg, 2006; 32: 1459–1463.
7. Kohnen, T., Allen, D., Boureau, C.: et al.: European Multicenter Study of the Acrysof ReSTOR Apodized Diffractive Intraocular Lens. Ophthalmology 2006; 113: 578–584.
8. Kyprianou, G., Langrová, H., Rozsíval, P., et al.: Kontrastní citlivost v refrakční chirurgii. Čes a Slov Oftal, 2009; 65: 68–70.
9. Souza, C.,E., Muccioli, C., Soriano, E.,S., et al.: Visual Performance of Acrysof ReSTOR Apodized Diffractive IOL: A Prospective Comparative Trail. Am J Ophthalmol, 2006; 141: 827.
10. Vingolo, E.,M., Grenga, E., M., Lacobelli, L., et al.: Visual acuity and contrast sensitivity: Acrysof ReSTOR apodized diffractive versus Acrysof SA60AT monofocal intraocular lenses. J Cataract Refract Surg, 2007; 33: 1244–1247.
11. Vries, de N., Webers, C., Montés-Micó, R, et al.: Long term follow-up of a multifocal apodized diffractive intraocular lens after cataract surgery. J Cataract Refract Surg, 2008; 34: 1476–1482.
Labels
OphthalmologyArticle was published in
Czech and Slovak Ophthalmology
2010 Issue 1
Most read in this issue
- Our Two – Years Experience with the Bevacizumab (Avastin) Treatment of the Age Related Macular Degeneration Wet Form
- The Influence of the Lens Capsule Mechanical Polishing to the Secondary Cataract Development
- Bacillus Cereus Keratitis – Case Report
- The Impact of Implantation of Intraocular Lenses with Negative Spherical Aberration on Contrast Sensitivity