Initial experience with the Vivasight ETView®double-lumen fibreoptic tube in thoracic surgery


Authors: Bicek Vladimír;  Krečmerová Martina;  Ungermannová Veronika;  Vymazal Tomáš
Authors‘ workplace: Klinika anesteziologie, resuscitace a intenzivní medicíny 2. LF UK ve FN v Motole
Published in: Anest. intenziv. Med., 25, 2014, č. 4, s. 274-276
Category: Anaesthesiology - Original Paper

Overview

Objective:
The double-lumen fibreoptic tube Vivasight ETView® is a new tool that allows lung isolation without the use of a fibrescope. The aim of the study was to evaluate the success and complication rates of its insertion in thoracic surgery procedures.

Design:
Prospective observational study (clinical audit).

Settings:
Department of Anaesthesiology and Intensive Care Medicine, University Hospital.

Material and methods:
We evaluated the success rate, speed of insertion, incidence of malposition and other complications associated with the insertion of a fibreoptic double-lumen tube in 65 patients scheduled for thoracic surgery requiring separation of the lungs.

Results:
The tube was successfully inserted in 64 of 65 patients (98.5%). One patient (1.5%) experienced tube dislocation requiring subsequent repositioning using a flexible bronchoscope. The average speed of insertion was 23.4 (± 4.67) seconds. We did not observe any significant differences in the insertion success rates between experienced anaesthesiologists and trainees.

Conclusion:
The double-lumen Vivasight ETView® fibreoptic cannula is a good tool for the separation of the lungs in thoracic anaesthesia. It is also suitable for educational purposes.

Keywords:
thoracic anaesthesia – double-lumen fibreoptic tube – Vivasight ETView® – insertion success rate


Sources

1. Seymour, A. H., Prasat B., McKenzie, R. J. Audit of double-lumen endobronchial intubation, Br. J. Anaesth., 2004, 93, p. 525–527.

2. Giglio, M., Oreste, D., Oreste, N. Usefulness of ETView TVT endotracheal tube for correct posi-tioning of bronchial blockers in left lobectomy: an easy and safe combination, Minerva Anestesiol., 2009, 75, p. 1–4.

3. Huitink, J. M., Koopman, E. M., Bouwman, R. A., Craenen, A., Verwoert, M., Krage, R., Visser, I. E., Erwteman, M., van Groeningen, D., Tinino, R., Schauer, A. Tracheal intubation with a camera embedded in the tube tip (Vivasight(™)). Anaesthesia., 2013, 68, p. 74–78.

4. Gaitini, L. A., Yanovski, B., Mustafa, S., Hagberg, C. A., Mora, P. C., Vaida S. J. A feasibility study using the VivaSight Single Lumen™ to intubate the trachea through the Fastrach laryn-geal mask airway: a preliminary report of 50 cases. Anesth. Analg., 2013, 116, p. 604–608.

Labels
Anaesthesiology, Resuscitation and Inten Intensive Care Medicine

Article was published in

Anaesthesiology and Intensive Care Medicine

Issue 4

2014 Issue 4

Most read in this issue
Topics Journals
Login
Forgotten password

Enter the email address that you registered with. We will send you instructions on how to set a new password.

Login

Don‘t have an account?  Create new account

#ADS_BOTTOM_SCRIPTS#