#PAGE_PARAMS# #ADS_HEAD_SCRIPTS# #MICRODATA#

THE EFFECT OF FRAME RATE AND CALIBRATION ON LUNG MONITORING WITH ELECTRICAL IMPEDANCE TOMOGRAPHY


Authors: Kristyna Koldova;  David Slajfercik
Published in: Lékař a technika - Clinician and Technology No. 4, 2019, 49, 107-111
Category:

Overview

Electrical impedance tomography (EIT) is a non-invasive method that records changes in tissue impedance in the selected cross-section of the body and is mainly used to monitor patient’s lung ventilation. The aim of this study was to analyze the effect of the pre-set frame rate and how calibration of the EIT system affects final record when monitoring lungs using EIT system PulmoVista 500 (Dräger Medical, Germany). Ten healthy male volunteers were measured with frame rates 10, 20, 30, 40 and 50 Hz for analysis, followed by calibrated and uncalibrated measurements. The subjects were breathing spontaneously in a horizontal supine position. In case of the frame rate analysis, data were statistically analyzed using ANOVA for repeated measures. Paired T-test was used for comparing data from calibrated and uncalibrated records. This study shows that the effect of both frame rate and calibration is negligible.

Keywords:

electrical impedance tomography (EIT) – EIT calibration – frame rate – lung monitoring


Sources
  1. Leonhardt S, Lachmann B. Electrical impedance tomography: the holy grail of ventilation and perfusion monitoring?. Intensive Care Medicine. 2012;38(12):1917–29.
  2. Teschner E, Imhoff M. Electrical impedance tomography: The realization of regional lung monitoring. Dräger Medical GmbH EIT Booklet, Germany, 2011.
  3. Brown BH. Electrical impedance tomography (EIT): a review. Journal of Medical Engineering & Technology. 2003;27(3):97–108. DOI: 10.1080/0309190021000059687
  4. Bodenstein M, David M, Markstaller K. Principles of electrical impedance tomography and its clinical application. Critical care medicine. 2009;37(2):713–24. DOI: 10.1097/CCM.0b013e3181958d2f
  5. Holder DS. Electrical impedance tomography: methods, history, and applications. Philadelphia: Institute of Physics Pub. 2005, xiii, 456 p. ISBN: 9781420034462.
  6. Buzkova K, Muller M, Rara A, Roubik K, Tyll T. Ultrasound detection of diaphragm position in the region for lung monitoring by electrical impedance tomography during laparoscopy. Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2018 Mar;162(1):43–6. DOI: 10.5507/bp.2018.005
Labels
Biomedicine
Topics Journals
Login
Forgotten password

Enter the email address that you registered with. We will send you instructions on how to set a new password.

Login

Don‘t have an account?  Create new account

#ADS_BOTTOM_SCRIPTS#